Ir arriba
Información del artículo en conferencia

Identification of parameters in a rogowski coil used for the measurement of partial discharges

G. Robles, M. Argueso, J. Sanz, R. Giannetti, B. Tellini

IEEE Instrumentation and Measurement Technology Conference - IMTC 2007, Varsovia (Polonia). 01-03 mayo 2007


Resumen:
The Rogowski coil is a simple and inexpensive device that can be adapted to measure high frequency currents such as those created by partial discharges in dielectrics. It has been demonstrated in previous works that the Rogowski coil can be self-integrating when used at high frequencies and can be tuned to measure partial discharges without the need of umbersome electronic devices. The aim of this paper is to explore the possibilities of obtaining better measurements simply changing the electric parameters and hence, the geometry of the coil. In the process, the inductance, mutual inductance, resistance and capacitance are written in function of geometric variables. Then, a bode plot simulates the effect of changing some of these variables and the results are commented focused on the pursued objective.


Palabras clave: Partial discharge measurements, Rogowski coil, instrument design, signal integration, voltage transducer.


DOI: DOI icon 10.1109/IMTC.2007.379175

Fecha de publicación: mayo 2007.



Cita:
Robles, G., Argueso, M., Sanz, J., Giannetti, R., Tellini, B., Identification of parameters in a rogowski coil used for the measurement of partial discharges, IEEE Instrumentation and Measurement Technology Conference - IMTC 2007, Varsovia (Polonia). 01-03 mayo 2007.


    Líneas de investigación:
  • *Instrumentación Electrónica

IIT-07-047A

pdf Solicitar el artículo completo a los autores