Go top
Conference paper information

Application of NIR laser diodes to pulse oximetry

S.M. López-Silva, R. Giannetti, M.L. Dotor, J.R. Sendra, J.P. Silveira, F. Briones

European Conference on Biomedical Optics - BIOS Europe 98, Stockholm (Sweden). 08-10 septiembre 1998


Summary:

A transmittance pulse oximeter based on near-infrared laser diodes for monitoring arterial blood hemoglobin oxygen saturation has been developed and tested. The measurement system consists of the optical sensor, sensor electronics, acquisition board and personal computer. The system has been tested in a two parts experimental study involving human volunteers. A calibration curve was derived and healthy volunteers were monitored under normal and apnoea conditions both with the proposed system and with a commercial pulse oximeter. The obtained results demonstrate the feasibility of using a sensor with laser diodes emitting at specific near infrared wavelengths for pulse oximetry.


Keywords: Hemoglobin oxygen saturation; Infrared detectors; Laser applications; Optical sensors; Oximeters


DOI: DOI icon 10.1117/12.336944

Publication date: September 1998.



Citation:
López-Silva, S.M., Giannetti, R., Dotor, M.L., Sendra, J.R., Silveira, J.P., Briones, F., Application of NIR laser diodes to pulse oximetry, European Conference on Biomedical Optics - BIOS Europe 98, Stockholm (Sweden). 08-10 September 1998.

IIT-98-085A